QA/QC Dissolved Oxygen

Sensor Information:

We currently deploy four types of sensors which measure Dissolved Oxygen. These are summarized in the following table:

Click sensor name for link to manufacturer website
SeaBird SBE 43
SeaBird SBE 37-SMP-IDO SeaBird SBE 63 Aandera Optode (3835)
Sensor Methodology Polarographic/Clark electrode type sensor Polarographic/Clark electrode type sensor Optical/Luminescent Optical/Luminescent
Accuracy +/-2% (Initial) +/-2% (Initial) +/-2% (Initial)  <8uM below 160uM & 5% above 160uM

Calibration Evaluation:
The dissolved oxygen sensors are calibrated by the manufacturer. Sensors are periodically recovered from the stations (where possible)  for calibration checks. The length of deployment between sensor evaluations will be a function of many factors including  the type of station and ease of access as well as sensor type.  When recovered, the sensors are placed in a well stirred and aerated tank and the readings are checked against each other and typically against one or more freshly calibrated sensors and a YSI optode.  For remote buoy deployments the sensors are checked prior to and following recovery. The results of these sensor cross-checks are used to determine when a sensor should be returned to the manufacturer for re-calibration.  The sensor check data are also used to calculate corrections for measured drift and the data quality is flagged accordingly. Details of data corrections will be provided in the metadata, may be described in a supporting document and can be found on the QA/QC Archive link at the bottom of this page.

 

Dissolved Oxygen QA/QC Archive

PDF or hyperlink Dates Description
SATURN-01    
data quality summary 8/3/10 - 1/31/17 evaluation/validation of calibration & sensor performance
s1-do01   Identifying and correcting for periods of insufficent flow to the sensor
SATURN-02    

data quality summary

2011 -2019 evaluation/validation of calibration & sensor performance
s2-do01   estimated salinity for 1m sensor deployment (2012)
SATURN-03    
data quality summary 6/13/10 - 7/15/20 evaluation/validation of calibration & sensor performance
  s3-do01 6/13/10 - 3/2/11 performance evalution; drift correction
s3_do02 12/18/12 - 5/31/13 performance evalution; drift correction
s3-do03 5/31/13 - 2/20/14 performance evalution; drift correction
s3-do04 2/20/14 - 11/8/14 performance evalution; drift correction
s3-do05 11/8/14 - 4/3/15 performance evalution; drift correction
s3-do06 4/4/15 - 10/30/15 performance evalution; drift correction
s3-do08 10/30/15 - 9/1/17 performance evalution; drift correction
s3-do09 4/19/19 - 7/13/20 performance evalution; drift correction
s3-do10 7/1/20 - 12/1/20 performance evalution; drift correction
SATURN-04    
data quality summary 9/23/09 - 7/1/20 evaluation/validation of calibration & sensor performance
s4-do01 4/8/13 - 2/20/14 performance evalution; drift correction
s4-do02 2/20/14 - 11/8/14 performance evalution; drift correction
s4-do03 11/8/14-2/23/16 performance evalution; drift correction
s4-do04 2/23/16 - 9/1/16 performance evalution; drift correction
s4_do05 4/14/18 - 7/1/20 performance evalution; drift correction
s4_do06 7/1/20 - 12/1/20 performance evalution; drift correction
SATURN-07    
data quality summary 12/4/12 - 5/1/19 evaluation/validation of calibration & sensor performance
s7-do01 1/15/15 - 2/1/17 performance evalution; drift correction
SATURN-09    
data quality summary 9/8/14 - 2/1/20 evaluation/validation of calibration & sensor performance
s9-do01 10/10/14 - 6/10/16 evaluation/validation of calibration
SATURN-10   evaluation/validation of calibration & sensor performance
data quality summary 8/28/16 - 12/17/16